Used regularly, TempTAB can help establish control over your sintering process. Logging your results will allow you to build a database of your furnace's thermal history to spot trends before they affect your zirconia.
Kit includes: TempTAB Snap-Cal offers precise and consistent measurements, featuring an ergonomic grip and a secure sample holder for enhanced accuracy. Also includes 1 sleeve of TempTAB 700 (25 Pieces) with a temperature range 1400 - 1750C / 2252 - 3092F.
How do I calibrate my furnace with TempTAB?
- Place a TempTAB in a sintering tray filled with beads and set it in your furnace
- Program your furnace to heat at 5°C/minute to1520°C with a 2-hour soak.
- After the heating cycle is complete, measure the diameter of TempTAB with a digital caliper.
- Using that measured dimension, look up its equivalent temperature in the appropriate temperature table.
- If the difference between the equivalent temperature and target temperature, 1520°C, is less than 5°C, then your furnace is calibrated.
- When the difference is over 5°C, adjust your target temperature by adding or subtracting that difference into the final programmed temperature and repeat the calibration run. If the TempTAB measures within the +/-5°C range , use that initial offset for future firings to reach your desired target temperature.
- If the initial difference is off by more than 25°C, consider replacing the thermocouple, or refer to your operating manual for help.
Dental Calibration Kit
SKU: P8 700 KIT
$400.00Price